If you have any questions or experience any problems, please call 0161 850 0759
For Diabetics 50 Strips
Blood volume : 2 μL
Measuring range : 10 mg/dL – 600 mg/dL
Sample types : Capillary blood. If the test strip is removed from the meter to apply blood: venous blood anticoagulated with lithium heparin or ammonium heparin or EDTA arterial blood and blood from neonates
Sample dosing : Blood application on to the middle of the test pad. The test strip (and its components) has the function to spread the blood above the test area – in a quick and hygienic way
Dosing options: Blood can be applied to both an already inserted test strip, and a tests trip outside the meter. This is helpful to minimize the risk of contamination when a meter has multiple users (e.g. in a hospital) and, in case of AST, to facilitate blood application
Re-dosing capability: Additional blood can be added to the test strip within 10 seconds after applying the first drop (applies for inside mode)
Visual color check: Plausibility check option with 5 color block on the test strip vial
Hematocrit range: 20 – 55 % (within-meter mode) 20 – 70 % (outside-meter mode)